custo guard holter mit den Klebeelektroden custo wing auf transparentem Hintergrund. Das Bild zeigt die richtige Anlage der Elektroden.Image Added



Technical specifications

The custo guard holter is supported from custo diagnostic version 4.6.x or custo diagnostic 5.0.x and higher.


Number of channels3

Sampling Frequency

(Depending on recording duration)

  • 533 Hz (24 h, high resolution)
  • 400 Hz (24 h, 48 h, 72 h)
  • 100 Hz (5 days, 7 days, unlimited duration)
Resolution

18 bit

3-dB bandwidth ECG amplifier
(depending on sampling frequency)
  • f0 Hz – 105 Hz (with a sampling frequency of 533 Hz)
  • 0 Hz – 80 Hz (with a sampling frequency of 400 Hz)
  • 0 Hz – 20 Hz (with a sampling frequency of 100 Hz)
Minimum detection threshold
(depending on sampling frequency
  • 10.46 μV (with a sampling frequency of 533 Hz)
  • 5.51 μV (with a sampling frequency of 400 Hz)
  • 5.51 μV (with a sampling frequency of 100 Hz)
ECG measurement range+/- 685.7 mV.
Radio frequency bandISM 2.4 GHz
Radio transmission power
  • 10 mW
Radio communicationFHSS Frequency Hopping Spread Spectrum
Running Time
  • internal rechargeable Battery approx. 10h
  • AAA-Batteries approx. 5h
Radio modulationGFSK

Radio transmission rate

1Mbit/s

Connection to PC

via custo guard base charging and communication unit
(USB) and Bluetooth

Power supply

Lithium polymer battery

Battery runtime)1

up to 168 h/7 days

Battery charging time

approx. 2 hours, charging current max. 130 mA

Status display

LED indicators for ECG quality and charging status

Dimensions

approx. 70 * 42 * 12 mm (L * W * H)

Weight

approx. 30 g

Ingress protection level

IP 65

Operating life

approx. 5 years (after this time has elapsed, the recording
duration of up to 168 h can no longer be guaranteed)

Operating conditions

  • Temperature: +5°C ... +45°C
  • Humidity: 10 ... 95 % rH
  • Air pressure: 700 ... 1060 hPa

Transport and storage conditions)2

Temperature: -20°C ... +45°C
Humidity: 10 ... 95 % rH
Air pressure: 700 ... 1060 hPa
Only for storage in a medical environment, under
consideration of the conditions mentioned here.

Classification

Device with internal power supply, Class IIa, Type BF

Applied standards

DIN EN ISO 13485, DIN EN ISO 14971, DIN EN ISO 20417, DIN EN ISO 15223-1, DIN EN 60601-1, DIN EN 60601-1-2, DIN-IEC 60601-1-6, DIN EN 62304, DIN EN 62366-1, DIN EN ISO 10993-1, DIN EN 60601-2-47, DIN EN 60601-1-11, DIN EN ISO 10993-5, DIN EN ISO 10993-10

...