Introduction

This article describes how to configure a exercise station with or without a training device to perform a training in Rehab. A exercise station with custo guard but without a training device serves purely ECG for monitoring and is preferred used for telemetry.

Configuration Exercise Station for Trainings with Training Device

Pre-Conditions

Configuration sequence

Preparation

Enter the Settings Page "Device Connection" with Steps 1-3

Image Added

Select the Number of Exercise Stations

Hint: If you want to use more than 8 exercise stations, 2 Monitors must be used.

Image Added

Select "Type of exercises"

The most common setting is "Free exercies with Smart-Cards". It means that the patient selection takes place individually via the context menu on the Exercise Station main page.
"With smart card" means: if the ergometer has a smart card reader, the selection of the patient takes place by inserting the smart card into the ergometer.
A detailed description of the types of exercise you can find here: https://www.customed.de/docs/x/H4CU.

Image Added

Configure the Exercise Station

Preparations

Each exercise station must be configured individually. To configure an exercise station you have to select the apropriate exercise station.
By click on the arrow button, its possible to change from station to station.

Starting is always from No. 1 and goes up to the maximum set number of exercise stations.

Image Added

Select the Training Device

Here you hate to assign a training device with integrated equipment like blood pressure measuring device, SpO2 sensor or and ECG amplifier to a exercise station.

  1. Click "Exercise device" in the list "Device list of exercise station"
  2. Select an exercise device (eg. Ergometer or Treadmill) from the list "Exercise devices" (see https://www.customed.de/docs/x/4gC also)
  3. If the training device has integrated options eg. chip card reader (smart card) or suction system (ECG device) you can enable or disable here these options.

Image Added

Select an ECG Device

If the training device have an integrated ECG device you can add here the ECG device to the exercise station.

  1. Click "ECG" in the list "Device list of exercise station"
  2. Select the apropriate device from the list "ECG device"

Image Added

Select the SpO2 Device

  1. Click "SpO2" in the list "Device list of exercise station"
  2. Select a device from the list "SpO2 device"

Image Added

Blood Pressure

If you use a training device with integrated blood pressure measuring device no action is here required.
If you want to use an external blood pressure measuring device (eg. Metronik BL-6, for example in combination with a treadmill) you have to configure this device here.

  1. Click "BP" in the list "Device list of exercise station"
  2. Select a device from the list "Blood pressure measuring device"

Image Added



Assigning the Training Device with guard

first Step

Image Added

All necessary COMRF sticks  are listed? f not, check the comRF  Installation, see https://www.customed.de/docs/x/IgKif yes, continue with the next step


Assigning guard

 Image Added

1: Click "ECG",  2: select "guard" and 3: Click "Save"


for fix Connection on the Trainingsplace:

Introduction: For a fast and effective work it is better to fix every single guard with a fixed training place.

The guards are sorted by the serial numbers and the lowest is assigned to training place 1 and for example the highest one to training place 16.


Assigning to guard
Anchor
Assigning to guard
Assigning to guard

next Step:

First, each guard is labeled with the training place number. This number makes it easier to assign to the training site later.

Image Added

Click "custo guard labeling" 

Image Added

1: Click "Sort" means: it is sorted from the lowest to the highest serial number 2: Enter the training place numbers 3: Click "Save"


next Step: fixed assignment of the guard to the training place

Image Added

1: activation the Setting "Allocate as fixed", 2 Enter "custo guard allocation"

Image Added


1: Button to select the found guards, here in the example the guard with the number 1 to training place 1.

  1. Configuration Trainingsplace for Training with Trainings Devices
  2. Telemetrie (pure ECG recording without training devices)
  1. Configuration Trainingsplace for Training with Trainings Devices

     1.1 Set the Basic Settings, see Device Setting for Stress Devices up to custo diagnostic 4.5.x

     1.2. Side Device Connection

Image Removed

     1.2. Assigning the Training Device (Equipment blood pressure, SPO2 and ECG amplifier)

Ergometer:

Image Removed

ecg:

Image Removed

SpO2:

Image Removed

Blood Pressure: 

does not have to be set, values are transferred automatically via the protocol

1.3. Assigning the Training Device with guard

1.3.1.1. Configuration comRF 

Install Driver:

Variante 1

via Setup Selection, see 

Image Removed

or Setup Command Line Parameters

1.3.1.2. Settings for comRF

Connect comRF to USB Port, and than configure the 1:1 Connection comRF to guard (as an Example, for 2 Trainingsplaces are needed 3 comRF Sticks)

Image Removed

1.3.2. Assigning guard

 Image Removed

for fix Connection on the Trainingsplace:

Image Removed

-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------


Telemetry

Conditions:
Telemetry means: There is only
2. Telemetrie (pure ECG recording without training devices)2.1. device control instead.

Trainingsplace without Trainings Device


1: Click "Selection Exercise"  2: select "No device" and Click "Save"

for Setting guard see Chapter Assigning to 2.2 Configure guard, see Section 1.3.2. Assigning guard