The article describes the display of the alarm type when the ECG is bad.
In rehab, the ECG signal quality is continuously determined.
It is possible to output a visual alarm when the ECG has reached the signal quality is defective or non-physiological.
The alarm cardio concept \ training place \ EcgQualityAlarmEnabled is deactivated by default and can only be set manually via the internal setting via the settings editor.
If the alarm is activated, a transparent red bar will be displayed in the online ECG on the overview page on the right. It automatically disappears when the ECG signal quality has improved.