To use this function, the suction level must be set to at least level 3 in custo diagnostic. With automatic suction power control, the suction level is increased until the system is airtight and the electrodes rest tightly enough against the skin of the patient. If the suction level is set manually, automatic suction level control is deactivated.
The maximum input voltage for the custo cardio 400 is 15 Volt DC. So the power supply from the custo cardio 200 can also be used. The power supply for the custo cardio 400 however is a medical power supply.
The custo cardio 400 will work on both devices. The Onyx Venus has to be used in 12 Volt DC output mode.
No, it is not possible to power the custo cardio 400 by USB ( this was also NOT possible for the custo cardio 200).
custo diagnostic 4.6 and custo diagnostic 5.0. (this however also applies to the custo cardio 300)
To use this function, the suction level must be set to at least level 3 in custo diagnostic. With automatic suction power control, the suction level is increased until the system is airtight and the electrodes rest tightly enough against the skin of the patient.
If the suction level is set manually, automatic suction level control is deactivated. Under level "3" the automatic suction control is not activated. ( It can be deactivated by pressing the "+" and "-" button on the cardio 400 simultaneously)
custo diagnostic can set 6 suction levels, from level 0 up till level 6 ( from 0 up till minus 280 mbar)
In manual mode you cannot set a lower value than step one ( -60 mbar) this to prevent that the user accidentally switches off the device and have all the electrodes drop off.
Blue: the device id connected using a wireless connection ( Bluetooth or WiFi)
turquoise: The device is connected using USB (USB-Modus)
Yellow:
Green: The automatic electrode dropoff is activated or the cardio 400 has been set into cleaning mode (<= 1min). The indication for the suction level has been used as countdown timer.
Orange: A key combination has been pressed e.g. to reset the cardio 400.
The cardio 400 can, in case of a warning, normally be used.
Press the ON- key for >= 3 Sec.
Pressure LED will light up Orange.
Depending on the settings in custo diagnostic will eitherĀ the cardio 400 activate the 10 seconds automatic resting ECG record or the ECG will be started and can be finished by pressing the ON key for another 3 seconds or stopping the record in custo diagnostic.
It is possible to configure custo diagnostic that it records 10 seconds of ECG (Setting : Autostart ) using start will record ECG till the next time the start button on the cardio 400 is pressed again. Or you can always stop the ECG in the custo diagnostic program.
You can update the firmware of the EAS01 circuit board by using the firmware update tool. This circuit board is responsible for the ECG amplifier. Inside the cardio 400 there is a second circuit board, the EAS02 the control board, this can only be updated in the factory.