The custo guard holter is supported from custo diagnostic version 4.6.x or custo diagnostic 5.0.x and higher.
| Number of channels | 3 |
Sampling Frequency (Depending on recording duration) |
|
| Resolution | 18 bit |
| 3-dB bandwidth ECG amplifier (depending on sampling frequency) |
|
| Minimum detection threshold (depending on sampling frequency |
|
| ECG measurement range | +/- 685.7 mV. |
| Radio frequency band | ISM 2.4 GHz |
| Radio transmission power |
|
| Radio communication | FHSS Frequency Hopping Spread Spectrum |
| Running Time |
|
| Radio modulation | GFSK |
Radio transmission rate | 1Mbit/s |
Connection to PC | via custo guard base charging and communication unit |
Power supply | Lithium polymer battery |
Battery runtime)1 | up to 168 h/7 days |
Battery charging time | approx. 2 hours, charging current max. 130 mA |
Status display | LED indicators for ECG quality and charging status |
Dimensions | approx. 70 * 42 * 12 mm (L * W * H) |
Weight | approx. 30 g |
Ingress protection level | IP 65 |
Operating life | approx. 5 years (after this time has elapsed, the recording |
Operating conditions |
|
Transport and storage conditions)2 | Temperature: -20°C ... +45°C |
Classification | Device with internal power supply, Class IIa, Type BF |
Applied standards | DIN EN ISO 13485, DIN EN ISO 14971, DIN EN ISO 20417, DIN EN ISO 15223-1, DIN EN 60601-1, DIN EN 60601-1-2, DIN-IEC 60601-1-6, DIN EN 62304, DIN EN 62366-1, DIN EN ISO 10993-1, DIN EN 60601-2-47, DIN EN 60601-1-11, DIN EN ISO 10993-5, DIN EN ISO 10993-10 |