Allgemeines zur HL7 Schnittstelle, custo diagnostic 5.6 or above, ACK,
nicht alle nachrichten lizenziert
short description - which shows the typically used fields / options.
General Data Used
Drübergehen DTM Format?
""
Key
Not listed fields are ignored
Default configuration - can be confgured in some extent.
Receiving / Sending seen from custo diagnostic perspective
optional - not reuqired, in some cases must be activated in custo diagnostic.
Field | o/r | Default Usage |
---|---|---|
MSH-3 | o | Sending Application. Incoming: Expected value can be configured (default empty) in custo diagnostic and works as a filter - meaning messages with deviant content in MSH-3 are ignored. Outgoing: Sending Application, e.g. "CUSTO" can be configured in custo diagnostic. |
MSH-4 | o | Sending Facility Incoming: Expected value can be configured in custo diagnostic and works as a filter - meaning messages with deviant content in MSH-4 are ignored. Outgoing: Sending Application, e.g. "CUSTO" can be configured in custo diagnostic. When a tenant system (german "Mandantensystem") is configured, this field contains the information about the tenant. See "Tenant System" below. |
MSH-5 | o | Receiving Application Incoming: Expected value can be configured (default empty) in custo diagnostic and works as a filter - meaning messages with deviant content in MSH-3 are ignored. Outgoing: Receiving Application, can be configured (default empty) in custo diagnostic e.g. "HIS". |
MSH-6 | o | Receiving Facility Incoming: Expected value can be configured in custo diagnostic and works as a filter - meaning messages with deviant content in MSH-3 are ignored. When a tenant system (german "Mandantensystem") is configured, this field should contain the information about the tenant. See "Tenant System" below. Outgoing: Receiving Facility, can be configured (default empty) in custo diagnostic e.g. "HIS". |
MSH-7 | r | Date/Time Of Message, see HL7 Standard. ##DTM## |
MSH-9 | r | Message Type. By default, custo diagnostic sends Messagetype^Event^Message-Structure, e.g. ORU^R01^ORU_R01. |
MSH-10 | r | Message Control ID, see HL7 Standard |
MSH-12 | r | Version ID, Outgoing: Custo diagnostic supports Version 2.3, 2.5 ##Verhalten bei eingehenden Nachrichten noch prüfen, wenn 2.7 2.9 oder 1.5 o.ä. drin steht## |
MSH-13 | r | Sequence Number, see HL7 Standard |
MSH-15 | ## | ##prüfen, sieh auch DS-840)## |
MSH-16 | ## | ##prüfen, sieh auch DS-840)## |
MSH-18 | o | Character Set ##zu prüfen## Incoming: see Standard Outgoing: By Default empty, ##Tatsächlich verwendetes CodeSet? ???? is used to encode message## |
### Hier bitte alles aus default Mapping übernehmen.###
z.B. PID-3, r, Patient-ID, see HL7 Standard. PID-3-1 is used to identify the patient in custo diagnostic. It must be unique in one custo diagnostic installation, otherwise an appropriate tenant system must be used (see tenant system below).
Required sind patienten name/vorname, geburtsdatum, Patienten-nummer, geschlecht?,
Field | o/r | Default Usage |
---|---|---|
PID-8 | r ? | Administrative Sex. Supported Values are F, M, ?? Other Values are interpeted as ... |
##bitte aus default mapping ergänzen## Ist wahrscheinlich alles optional
Field | o/r | Default Usage |
---|---|---|
PV1 | r | Visit-ID. The visit id is used to identify the visit in custo diagnostic It must be unique in one custo diagnostic installation, otherwise an appropriate tenant system must be used (see tenant system below). |
PV1-44 | o | Admission Date: custo diagnostic can store this information, even it is not shown in the user interface. Its done for future usage. |
PV1-45 | o | Discharge Date/Time: custo diagnostic can store the discharge date/time to filter out discharged patients in patient search lists. It can be defined, how custo diagnostic interpretes an empty discharge date/time - either als missing, unknown information, or as the information, that the discharge date should be deleted in custo diagnostic. |
## bitte hier alles aus default mapping nehmen##
Field | o/r | Default Usage |
---|---|---|
OBR-1 / OBR-5 | r | The combination of OBR-1 and OBR-5 can be configured as order control: |
OBR-2 | r | Placer ID |
OBR-3 | o | Sending: |
OBR-4 | r | OBR-4 contains information about the type of observation which should by executed. OBR-4-1 is a coded value which must be defined. OBR-4-2 when uutgoing, OBR-4-2 is shown to the user, e.g. "Resting ECG" ##bitte prüfen - when OBR-4-2 in eingehender Nachricht abweichen zur Konfiguration in "Untersuchung" - welcher Text wird dabei angezeigt - der aus der Konfig, oder der aus der Nachricht? |
custo diagnostic receives orders and show them as a worklist to the users. One order is (by default) identified by the Placer Number, and may contain serveral observation requests, identified by Alternate Order ID.