Lode bicycle ergometers and treadmills with integrated blood pressure device need to have an external QRS trigger to work properly ( the range with ergometers involved is listed below)

custo med GmbH offers a new trigger cable that can be used to trigger the build-in  blood pressure devices in the latest models of the Lode ergometers and treadmills.

This trigger cable can be offered for the cardio 1xx serie and the cardio 200


Note: The standard trigger cable with a BNC connector cannot be used for these ergometers and treadmills.

 

The new option is named “Realtime-QRS-Trigger-Module” and can be ordered by using part number 12149/LODE ( this exists out of a cable and a software license)

 

The trigger cable that can be used to trigger the internal blood pressure device, must be used for the ergometers and treadmill out of the following range:

 

The trigger cable has a DB-9 connector on one end and a 2,5 mm2 on the other end. The cable must be connected to the PC  using an RS-232 port. A serial to USB converter might cause a delay in the trigger pulse which will affect correct blood pressure measurements, We strongly recommend NOT to use USB to serial converters because of that.