The article describes, a easy quality check from guard and belt. The condition before is: https://www.customed.de/docs/x/8gC => Chapter "Assigning the Training Device with guard"
Belt connect to guard
- Both electrodes are connected to each other, if necessary moisten contact surfaces with saline solution
- Short circuit occurs
- Zero line in custo diagnostic is visible after 2 seconds at the latest
- Movement of the guards and the belt result in no change of the zero line, this runs rigidly
Error:
Changes the zero line and disturbances occur:
- Belt swap and create short circuit, fault is gone, cause is due to the Belt
- Reverse guard swap, cause may be due to guard
=> To be sure: if the belt / guard is broken, use a new one and test it.
- Belt is connected to guard
- Both electrodes are connected to each other, if necessary moisten contact surfaces with saline solution
- Short circuit occurs
- Zero line in custo diagnostic is visible after 2 seconds at the latest
- Movement of the guards and the belt result in no change of the zero line, this runs rigidly
- 3rd electrode is additionally connected
- Zero line in custo diagnostic is visible on all 3 channels
- Movement of the guards and the belt result in no change of the zero line, this runs rigidly
Error:
Changes the zero line and disturbances occur:
- Belt swap and short circuit, fault is gone, cause may be due to the Belt
- Reverse guard swap, cause may be due to guard
=>To be sure: if the belt / guard is broken, use a new one and test it.