Introduction:
This article describes how to configure a Trainings Place with and without a training device.
A trainings place without training device with guard, serves purely for ECG monitoring and is referred to as telemetry.
Configuration Trainingsplace for Training with Trainings Devices
Conditions:
- Key-Options, see https://www.customed.de/docs/x/DQK
- set the Basic Settings from Device Trademill/Ergometer in custo diagnostic see https://www.customed.de/docs/x/4gC
- set the comRF Connection, see https://www.customed.de/docs/x/IgK
Configuration sequence:
Enter the Page Device Connection with Steps 1-3
next Step, Selcetion of Trainingsplaces
if there are more than 8 places, 2 Monitors must be used.
next Step, set "Type of exercises"
The most common setting is "Free exercies with Smart-Cards". It means that the patient selection takes place individually via the context menu on the Trainings Place main page.
With smart card means: if the ergometer has a chip card reader, the selection of the patient takes place by inserting the smart Card.
A detailed description of the types of exercise is here: https://www.customed.de/docs/x/H4CU
next Step
Selection of Trainingsplaces
Each training place must be configured separately.
By click on the button see arrow, it is possible to change from place to place.
Starting is always No. 1 and extends to the maximum set number, see Selection of Trainingsplaces
next Step
Assigning the Training Device with Equipment blood pressure, SPO2 and ECG amplifier from an ergometer to a training place
Selection of the ergometer
1: Click "Selection Exercise" 2: Selection of ergometers from the list of devices (see https://www.customed.de/docs/x/4gC) and Click "Save"
ecg:
1: Click "ECG" and 2: select "Exercice Device" and Click "Save"
SpO2:
1: Click "SPO2" and 2: select "Exercice Device" and Click "Save"
Blood Pressure:
does not have to be set, values are transferred automatically via the protocol
Assigning the Training Device with guard
first Step
All necessary COMRF sticks are listed? f not, check the comRF Installation, see https://www.customed.de/docs/x/IgK, if yes, continue with the next step
Assigning guard
1: Click "ECG", 2: select "guard" and 3: Click "Save"
for fix Connection on the Trainingsplace:
Introduction: For a fast and effective work it is better to fix every single guard with a fixed training place.
The guards are sorted by the serial numbers and the lowest is assigned to training place 1 and for example the highest one to training place 16.
Assigning to guard
next Step:
First, each guard is labeled with the training place number. This number makes it easier to assign to the training site later.
Click "custo guard labeling"
1: Click "Sort" means: it is sorted from the lowest to the highest serial number, 2: Enter the training place numbers 3: Click "Save"
next Step: fixed assignment of the guard to the training place
1: activation the Setting "Allocate as fixed", 2 Enter "custo guard allocation"
1: Button to select the found guards, here in the example the guard with the number 1 to training place 1.
-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Telemetrie
Conditions:
Telemetry means: There is only ECG recording without training device control instead.
Trainingsplace without Trainings Device
1: Click "Selection Exercise" 2: select "No device" and Click "Save"
for Setting guard see Chapter Assigning to guard